SLICE alignment workflow

Purpose

Use the SLICE alignment workflow to adjust the position of the light sheets so that they intersect with the focal plane of the detection objective. Proper alignment is necessary to achieve sharp in focus images.

Alignment enables the ability to get crisp focus at a single Z plane. Proper alignment in conjunction with a correctly adjusted offset setting enables crisp focus at different Z planes. You will need to adjust the offset value as you navigate around a tissue sample. Refractive Index (RI) compensation mapping enables you to record the proper offset position at different locations of the tissue sample and an RI compensation map is required for acquiring a Tissue Scan. However, if you are just navigating around the tissue live, you can manually adjust the offset in the SLICE Light sheet panel as you go.

Procedure

Before you start

  • Remove the oil reservoir drip tray from the microscope.

  • The alignment pattern is a series of lines etched into the alignment prism mirror in random orientations. This may make the mirror appear filmy or dirty upon visual inspection before being attached to the microscope stage.

  • When using the joystick, adjust the slider to lower the step size for finer stage movements. This will enable more exact positioning of the mirror alignment patterns and light sheets in the field of view as needed.

  • As you move through the workflow, the light sheet controls will automatically adjust as needed for each step.

Click SLICE Alignment on the Advanced ribbon to get started.

If the Neutral density filter was not automatically selected upon launching the workflow, manually select it in the SLICE Light Sheet panel, Emission Filter dropdown menu.
Follow this procedure to configure the Neutral density filter to be automatically selected when the SLICE alignment workflow starts.

SLICE alignment workflow steps

  1. Attach Mirror

    Attach the alignment prism mirror to the microscope stage. The stage will automatically move to the load position to attach the mirror.

    The prism mirror is designed to only attach in the correct orientation.

  2. Start left sheet alignment

    Turn on the left light sheet and bring the alignment pattern on the mirror into the field of view.

  3. Find center in X

    Find the center of the alignment pattern on the alignment prism mirror.

  4. Translate Detection Objective

    Physically adjust the position of the detection objective so that its focal plane intersects with the left light sheet.

  5. Adjust left sheet Y center

    Adjust the Center setting in the Left Sheet Controls section of SLICE Light sheet panel to center the light sheet vertically in the field of view. When entering this step, the height of the light sheet will automatically adjust to the lowest setting = 1.

  6. Translate Left Illumination Objective

    Focus the left light sheet by adjusting the position of the left illumination objective. Physically turn the knob below the left illumination objective underneath the left side of the SLICE unit until the light sheet is as thin as possible and in sharp focus. The knob will turn forward and backward.

  7. Balance illumination left

    Balance the left sheet illumination in the field of view.

  8. Start right sheet alignment

    Adjust the right light sheet so that it intersects with the focal plane of the detection objective.

  9. Adjust right sheet Y center

    Adjust the Center setting in the Right Sheet Controls section of SLICE Light sheet panel to center the light sheet vertically in the field of view. When entering this step, the height of the light sheet will automatically adjust to the lowest setting = 1.

  10. Translate Right Illumination Objective

    Focus the right light sheet by adjusting the position of the right illumination objective. Physically turn the knob below the right illumination objective underneath the right side of the SLICE unit until the light sheet is as thin as possible and in sharp focus. The knob will turn forward and backward.

  11. Balance illumination right

    Balance the right sheet illumination in the field of view.

  12. Remove mirror

    The stage will automatically move to load position and you can remove the alignment prism mirror from the microscope stage.

    When finished, click Close SLICE align workflow