You are here: Calibrating the system > Troubleshooting > Verifying camera to stage alignment

Camera/stage alignment

 

  1. Verify that the camera is oriented in the proper direction: What appears through the eyepieces should be identical in orientation to what appears on the screen.

    Example with the MBF calibration grid slide

    If the small grid is to the left of the large grid when looking through the eyepieces, then it should appear the same way on the computer screen. If it appears backwards, you need to rotate the camera 180°.

  2. Place a reference point on the screen (anywhere is fine).
  3. Click the Display Grid icon ; a white grid appears on the screen, displaying horizontal and vertical reference lines.
  4. For a preliminary alignment, at a low objective (2X or 5X), locate the large grid on the slide (250 µm) and align it closely to the grid on the screen .
  5. Continue in Joy Free mode at a higher objective (10X or 20X).
  6. Move the dust particle to the top of a white grid line on one side of the screen.
  7. Continue moving the dust particle across the screen and adjusting for correction if needed.
  8. Move to a higher objective and repeat steps 5–7 for a more fine-tuned alignment.
  9. Click the Grid icon again to turn it off.

 

Once the alignment is done, tighten the C-mount securely into place. As you tighten the screws, the C-mount may move slightly. Make sure that you anticipate that movement as you align the camera.

 

See Calibration—Troubleshooting


Microlucida 3 | MBF Bioscience Support Center | Downloads