Creating a new lens based on a calibration grid slide

Procedure

  1. Acquire an image of the MBF calibration grid slide using the same objective that you plan to use in your study.
  2. Load the grid(s) image into Microlucida. The Image Scaling dialog box opens .
  3. Open the Image Organizer:
  4. Make sure only the relevant grid image is visible:
  5. Click Tools>Create New Lens (Tools>Define New Lens in versions 10 and older).

    If you are calibrating from an image with a scale bar instead of the MBF calibration grid slide, check the Force Square checkbox.

  6. If defining multiple lenses from calibration grid images, hide the imageclick on the icon to deselect the image in the Image Organizer for the lens that has just been defined and repeat this process for the other lenses.
  7. Use the Quick Measurements Functions to verify that the calibration is correct (Quick Measurement).

 

See Calibrating your System